Liquiline System CA80SI
Operation
Hauser
97
Menu/Setup/Additional functions/Time switches/ Time switch 1 ... 8
Function
Options
Info
Function
Selection
• On
• Off
Factory setting
Off
Switches the function on or off
Start date
01.01.2000 to 31.12.2099
Format
DD.MM.YYYY
‣
Enter start date
Start time
00:00:00 to 23:59:59
Format
hh.mm.ss
‣
Enter start time
Signal duration
00:00:03 to 2400:00:00
Format
hh.mm.ss
Duration of the high signal level at the start of a
cycle
Period
00:00:03 to 2400:00:00
Format
hh.mm.ss
Duration of a cycle
Signal
Display only
Current process value of the time switch
Next signal date
Display only
Date of the next signal
Next signal time
Display only
Time of the next signal
Example 1: Time-based set point for a temperature controller
The temperature should increase to 21 °C from 08:00 every day, and then to 23 °C for 5
hours from 12:00. The temperature should be controlled to fall back down to 18 °C from
17.00. Two time switches are defined for this purpose which are used in a
MF1: Formula
mathematical function. Using the mathematical function, an analog temperature set point
is thus available for a controller.
T
[° C]
t [h]
18
21
23
8
12
17
0
T2
T1
A0041704
71
Time-based temperature control
1. Program
Time switch 1
(T1):
•
Start date
= 01.01.2020
•
Start time
= 08:00:00
•
Signal duration
= 09:00:00
•
Period
= 24:00:00
2. Define
Time switch 2
(T2):