Operation
Liquiline System CA80SI
60
Hauser
• Logbook name
Assign a name. Example: "01".
• Source of data
Select a data source. Example: Signal of binary input 1.
• Measured value
Select the measured value to be recorded. Example: pH value.
• Scan time
Specify the time interval between two logbook entries.
• Data logbook
Activate the logbook: specify the data storage method.
2. ../
Finished
: Perform the action.
The device shows the new logbook in the list of data logbooks.
3. Select data logbook "01".
Additional display:
Log time left
.
4. Only in the case of
Fill up buffer
:
Decide to set
Overflow warning
:
On
or
Off
.
On
: The device displays a diagnostic message in the event of memory overflow.
5.
Line plotter
submenu: Specify the type of graphic representation.
10.2.5 Extended setup
Diagnostics settings
The list of diagnostic messages displayed depends on the path selected. There are device-
specific messages, and messages that depend on what sensor is connected.
Menu/Setup/(General settings or Inputs<Sensor channel>)/Extended setup/Diagnostics settings/Diag.
behavior
Function
Options
Info
List of diagnostic messages
‣
Select the message to be changed. Only then
is it possible to make the settings for this
message.
Diag. code
Read only
Diagnostic message
Selection
• On
• Off
Factory setting
Depends on the message
Deactivate the diagnostic message or activate it
again.
Deactivating means:
• No error message in the measuring mode
• No failure current at the current output
Failure current
Selection
• On
• Off
Factory setting
Depends on the message
‣
Decide whether a failure current should be
output at the current output if the diagnostic
message display is activated.
In the event of general device errors, the
failure current is output at all the current
outputs. In the event of channel-specific
errors, the failure current is only output at
the assigned current output.
Status signal
Selection
• Maintenance (M)
• Out of specification (S)
• Function check (C)
• Failure (F)
Factory setting
Depends on the message
The messages are divided into different error
categories in accordance with NAMUR NE 107.
‣
Decide whether a status signal assignment
should be changed for the application.