Liquiline System CA80SI
Technical data
Hauser
163
Web server
The web server enables full access to the device configuration, measured values, diagnostic
messages, logbooks and service data via standard WiFi/WLAN/LAN/GSM or 3G routers
with a user-defined IP address.
TCP port
80
Supported features
• Remote-controlled device configuration(1 session)
• Save/restore device configuration (via SD card)
• Logbook export (file formats: CSV, FDM)
• Access to web server via DTM or Internet Explorer
• Login
• Web server can be switched off
15.6 Power supply
Supply voltage
• 100 to 120 V AC / 200 to 240 V AC
• 50 or 60 Hz
Fieldbus connection
Supply voltage: not applicable
Power consumption
For a sample flow rate of 80 ml/min (2.7 fl oz/min), a continuous measuring interval
(10 minutes), a sample temperature of 25 °C (77 °F), an ambient temperature of 25 °C
(77 °F) and a device with a supply voltage of 230 V:
• Typically 60 VA
• Max. 1250 VA
Cable entries
• 4 x bores for M16, G3/8, NPT3/8", Memosens connection
• 4 x bores for M20, G1/2, NPT1/2"
Cable specification
Cable gland
Permitted cable diameter
M16x1.5 mm
4 to 8 mm (0.16 to 0.32")
M12x1.5 mm (for order version M12 socket for Memosens sensors)
2 to 5 mm (0.08 to 0.20")
M20x1.5 mm
6 to 12 mm (0.24 to 0.48")
NPT³/₈"
4 to 8 mm (0.16 to 0.32")
G³/₈
4 to 8 mm (0.16 to 0.32")
NPT½"
6 to 12 mm (0.24 to 0.48")
G½
7 to 12 mm (0.28 to 0.48")
Cable glands mounted at the factory are tightened with 2 Nm.