Liquiline System CA80SI
Maintenance
Hauser
145
Replacing the pump hose
The replacement of the pump hose is not part of routine maintenance activities. Only
replace the pump hose when this is necessary.
1. Open the analyzer.
2. Stop the sample supply.
3. Rinse the reagent and standard channels.→ 140
There should not be any more liquid left in the hoses.
4. Remove the bottles and the bottle tray.
5. Press the soft key
MODE
and select
Manual mode
.
The display shows
Current mode
-
Manual
.
Wait until all operations have stopped. Ongoing operations can be stopped under
Menu
/
Operation
/
Manual operation
.
6. Open the bayonet lock of the peristaltic pump.
7. Replace the hose and, if necessary, replace the pump head.
8. Close the bayonet lock of the peristaltic pump.
9. Make sure all the hoses and connectors are seated correctly.
10. Place the bottle tray back into the housing.
12.2.7 Replacing the filter mats
You require the following parts:
Filter mats (part of the CAV maintenance kit)
A0045364
76
Fan screen on the bottom
1. Open both fan screens and remove them.
The filter mats are located under the screens.
2. Remove the used filter mats and replace them with new ones from the maintenance
kit.
3. Fit the fan screens back on.
4. Reset the operating hours counter for the mats:
Menu
/
Operation
/
Operating time
filter mats
/
Reset
.