System integration
Liquiline System CA80SI
36
Hauser
7
System integration
7.1
Overview of device description files
7.1.1
Device identification
→ 160, protocol-specific data
7.1.2
Source for device description and device master files
Downloading device drivers
.
A dropdown list is displayed.
2. Select: device driver.
3. Narrow your search by specifying the driver type, product root and communication
protocol.
4. Search.
5. Click the suitable entry from the list of hits.
You receive the download links including an extended description.
7.2
Integrating the measuring device into the system
7.2.1
Web server
Versions without fieldbus: An activation code is required for the web server.
Connection
‣
Connect the communication cable of the computer to the Ethernet port of the BASE2
module.
BASE2
Web Browser
A0044872
54
Web server/Ethernet connection
Establishing the data connection
All versions with the exception of PROFINET:
To ensure that your device has a valid IP address, you must disable the
DHCP
parameter in
the Ethernet settings. (
Menu
/
Setup
/
General settings
/
Extended setup
/
Ethernet
/
Settings
)
You can assign the IP address manually in the same menu (for point-to-point
connections).