Liquiline System CA80SI
Maintenance
Hauser
143
2.
1
Push the cuvette with the magnetic stir bar and cover into the holder. Ensure that the
tab (1) snaps into place.
3.
Connect the outlet hose "W" again and secure the cover again.
12.2.4 Replacing reagents
Removing bottles
1. Stop the sample supply.
2. If you have not already done so,
set the manual mode.
3. Remove the bottles you wish to replace by removing the hoses at the valves. Use a
paper towel to catch any reagents that may escape.
4. Select
Menu
/
Operation
/
Maintenance
/
Bottle change mode
/
Bottle removal
/
Bottle selection
.
5. Select the bottles that you want to remove and press
OK
to confirm.
6. Select
Confirm bottles removed
.
Replacing bottles
1. Replace the removed bottles with fresh reagent .
Reagents must be prepared according to the instructions for mixing a reagent.
2. Reconnect the hoses to the valves.
3. Select
Menu
/
Operation
/
Maintenance
/
Bottle change mode
/
Bottle insertion
/
Bottle selection
.
4. Select all the bottles that you have replaced and press
OK
to confirm.
5. Select
Confirm bottles inserted
.