Liquiline System CA80SI
Operation
Hauser
71
Behavior after power failure and other settings
Menu/Setup/Analyzer/Extended setup
Function
Options
Info
After power failure
Selection
• Last mode
• Manual mode
Factory setting
Last mode
Setting specifying how the analyzer should
behave after a power failure, or when the power
is back on.
Last mode
: The analyzer remains in the mode
that was last set. Example: The automatic mode
was set. The analyzer continues after
initialization and after any samples have been
discarded.
Manual mode
: The analyzer switches to the
manual mode and waits for user action.
Timeout no sample
Selection
0:00 to 30:00 mm:ss
Factory setting
1:00 (MM:SS)
Maximum time the device waits for sufficient
sample flow
10.3.2 Measurement
Menu/Setup/Analyzer/Measurement
Function
Options
Info
Start condition
Selection
• Immediate
• Date/time
Factory setting
Immediate
Immediate
: The analyzer immediately starts
with the measurement cycle once the system
changes to the automatic mode.
Date/time
: The analyzer starts with the
measurement cycle at the set date/time.
Date
01.01.1970 to 07.02.2106
Functions only available for:
Start condition
=
Date/time
Time
00:00:00 to 23:59:59
Measuring interval
0:10 to 24:00 (
hh:mm
)
Factory setting
0:20
Function only available for:
Start condition
=
Immediate
or
Date/time
For configuring the time interval of the
measurement
Sequence of measurements
Selection
• SP1
• SP2
• SP3
• SP4
• SP5
• SP6
• Pause
Defines the sequence of channels in automatic
mode.
Is displayed only in the case of a two-, four- or
six-channel device.
•
SP1
: Indicate the number of consecutive
measurements for channel
SP1
•
SP2
: Indicate the number of consecutive
measurements for channel
SP2
•
SP3
: Indicate the number of consecutive
measurements for channel
SP3
•
SP4
: Indicate the number of consecutive
measurements for channel
SP4
•
SP5
: Indicate the number of consecutive
measurements for channel
SP5
•
SP6
: Indicate the number of consecutive
measurements for channel
SP6
•
Pause
: If the
Pause
channel is selected, the
analyzer does not perform the measurement.
You can add, delete or save rows in the table
using the
INSERT
,
DEL
and
SAVE
soft keys.
Signal delay
0 to 600 s
Factory setting
0 s
Measurement active
: You can delay the process
of outputting the signal during an active
measurement by the defined time. Measuring is
paused for the duration of the signal delay time.