Liquiline System CA80SI
Technical data
Hauser
165
Sample requirement
> 140 ml (4.73 fl oz)/measurement, depending on flow
Reagent requirement
• 200 µl per reagent and measurement
for measuring intervals ≥ 15 minutes
• Given a measuring interval of 20 min, one set of reagents (1000 ml) lasts for approx. 65
days
• Given a measuring interval of 12 min, one set of reagents (2500 ml) lasts for approx.
100 days
Standard requirements
• Approx. 140 ml (4.73 fl.oz) / calibration
• With a 2.5 l (84.5 fl.oz) standard bottle and calibration interval of 96 h approx. 65 days
• With a 5 l (169 fl.oz) standard bottle and calibration interval of 96 h approx. 130 days
Calibration interval
1 to 99 days, depending on application and ambient conditions
Maintenance interval
Every 3 to 6 months, depending on the application
Maintenance effort
Weekly: visual inspection
15.8 Environment
Ambient temperature
range
+5 to +40 °C (41 to 104 °F)
Storage temperature
–20 to 60 °C (–4 to 140 °F)
Relative humidity
10 to 95 %, non-condensing
Degree of protection
IP55 (cabinet, analyzer stand), TYPE 3R (cabinet, analyzer stand)
Electromagnetic
compatibility
Interference emission and interference immunity as per EN 61326-1:2013, Class A for
Industry
Electrical safety
According to EN/IEC 61010-1:2010, Class I equipment
Low voltage: overvoltage category II
For installations up to 2000 m (6500 ft) above MSL
Pollution degree
Pollution level 2
15.9 Process
Sample temperature
5 to 45 °C (41 to 113 °F)
5)
We recommend the large reagent set 3 x 2.5 l + 1 x 5 l for measuring intervals < 15 minutes. We recommend the small reagent set 3 x 1 l + 1 x
2.5 l
6)
Sufficient mains quality is required to operate the product as intended.