Liquiline System CA80SI
Maintenance
Hauser
139
Interval
Maintenance work
12 months
‣
Replace the cuvette.
‣
Clean the magnetic stir bar.
‣
Check hose of hose pump for wear, and replace if necessary.
‣
Replace hoses.
‣
Replace filter mats.
If required
‣
Clean the housing.
‣
Clean the cuvette.
‣
Replace the cuvette.
‣
Clean the magnetic stir bar.
‣
Rinse the reagent and standard channels.
‣
Replace the hose connector.
‣
Clean the filter (backwash).
12.2 Maintenance tasks
12.2.1 Software menu Maintenance
Menu/Operation/Maintenance
• Bottle replacement
If level monitoring is activated: specify when bottles are replaced.
• Dispenser replacement
Operating time counter, can be reset here. Replace dispensers: → 146
• Filter mats replacement
Operating time counter, can be reset here. Replace filter mats: → 145
• Decommissioning
12.2.2 Setting the manual mode
Initial situation: The device is still in the automatic mode and you want to perform
maintenance work.
1. Press the
MODE
soft key.
The current mode is displayed:
Automatic
2. Select an action:
Manual mode
The device changes to the manual mode.
3. Wait until all operations have stopped.
4.
Alternatively:
Stop actions currently running under:
Menu
/
Operation
/
Manual operation
.
12.2.3 Cleaning
L
CAUTION
Risk of injury from reagents escaping
‣
Before replacing, always rinse the reagent and standard channels first.
Cleaning the housing
‣
Clean the front of the housing using commercially available cleaning agents only.
The front of the housing is resistant to the following in accordance with DIN 42 115:
• Ethanol (for a short time)
• Diluted acids (max. 2% HCl)
• Diluted bases (max. 3% NaOH)
• Soap-based household cleaning agents