Commissioning
Liquiline System CA80SI
52
Hauser
9.6
Starting the measurement
Pay attention to the following, particularly if measuring very low concentrations of
silica:
• The measurement results can display an initial drift. This can be caused by the possible
contamination of the components carrying sample.
• For this reason it is advisable to rinse the sample-bearing pipes for several hours with
continuous measurements prior to performing a calibration.
• The stability of the calibration factors can be checked by repeating the calibration
manually.
1. Select start condition
Immediate
under
Menu
/
Setup
/
Analyzer
/
Measurement
/
Start condition
/
Immediate
. The analyzer starts immediately with the measurement
cycle once the system changes to automatic mode.
2. If necessary, adjust the measuring interval under
Menu
/
Setup
/
Analyzer
/
Measurement
/
Measuring interval
.
3. If necessary, adjust the calibration interval under
Menu
/
Setup
/
Analyzer
/
Calibration
/
Calibration interval
.
4. If necessary, adjust the sequence of sample channels under
Menu
/
Setup
/
Analyzer
/
Measurement
/
Measuring interval
/
Sequence of measurements
.
5. Start automatic mode: Press the
MODE
and select
Start automatic mode
must be
selected.
The display shows
Current mode
-
Automatic
.
‣
Re-attach the cover in front of the cuvette assembly.