Liquiline System CA80SI
Technical data
Hauser
159
Signal on alarm
Adjustable, as per NAMUR Recommendation NE 43
• In measuring range 0 to 20 mA:
Failure current from 0 to 23 mA
• In measuring range 4 to 20 mA:
Failure current from 2.4 to 23 mA
• Factory setting for failure current for both measuring ranges:
21.5 mA
Load
Max. 500 Ω
Transmission behavior
Linear
15.3 Current outputs, active
Span
0 to 23 mA
Signal characteristic
Linear
Electrical specification
Output voltage
Max. 24 V
Test voltage
500 V
Cable specification
Cable type
Recommended: shielded cable
Cable specification
Max. 2.5 mm
2
(14 AWG)
15.4 Relay outputs
Electrical specification
Relay types
• 1 single-pin changeover contact (alarm relay)
• 2 or 4 single-pin changeover contacts (optional with extension modules)
Maximum load
• Alarm relay: 0.5 A
• All other relays: 2.0 A
Relay switching capacity
Base module (Alarm relay)
Switching voltage
Load (max.)
Switching cycles (min.)
230 V AC, cosΦ = 0.8 to 1
0.1 A
700,000
0.5 A
450,000
115 V AC, cosΦ = 0.8 to 1
0.1 A
1,000,000
0.5 A
650,000
24 V DC, L/R = 0 to 1 ms
0.1 A
500,000
0.5 A
350,000