Maintenance
Liquiline System CA80SI
144
Hauser
6. Following the replacement, a zero point calibration must be performed and the
calibration factor must be redetermined:
Menu
/
Operation
/
Manual operation
/
Determine zero point
and /
Determine calibration factor
.
7. Go back to the
MODE
/
Continue automatic mode
or
MODE
/
Start automatic mode
to start the normal measurement.
12.2.5 Performing zero point calibration
Replacing standard calibration solution with zero standard
1. If you have not already done so,
set the manual mode.
2. Replace the standard with zero standard. Use either an additional hose in order to
reconnect the standard calibration solution, or allow the hose to drain and dry the
hose well.
3. Select
Menu
/
Operation
/
Maintenance
/
Bottle change mode
/
Bottle insertion
/
Bottle selection
.
4. Select
Standard S1
and press
OK
to confirm.
5. Select
Confirm bottles inserted
.
Determining the zero point
1. Call up:
Menu
/
Operation
/
Manual operation
/
Determine zero point
.
2. Carry out zero point calibration.
3. Following successful calibration, the following message appears: "Do you want to
accept the calibration data for adjustment?". Press
OK
to confirm.
Re-deploying the standard calibration solution
‣
Replace the zero standard with the standard.
Recommendation: Afterwards, perform a manual calibration with the standard
solution (
Determine calibration factor
).
Alternatively, the zero point and calibration factor can be determined in one step
(
Determine zero point and calibr. factor
).
12.2.6 Replacing the hoses
1. Rinse the reagent and standard channels. → 140
2. Remove the bottles and the bottle tray.
3. Replace the hoses with hoses of the same diameter and length. Attach hose labels to
the new hoses.
4. Place the bottle tray back into the housing.
5. Connect the bottles to the appropriate covers and hoses.
6. Select
Menu
/
Operation
/
Maintenance
/
Bottle change mode
/
Bottle insertion
/
Bottle selection
.
7. Select all the bottles and press
OK
to confirm.
8. Select
Confirm bottles inserted
.
9. Press the
MODE
soft key and select
Continue automatic mode
or
Start automatic
mode
.