Maintenance
Liquiline System CA80SI
146
Hauser
12.2.8 Replacing dispensers
You require the following parts:
• From the CAV800 maintenance kit:
10 ml dispensers with adapter
• 1 pair of gloves that are resistant to the reagents used
Preparing to replace dispensers
1. Rinse the reagent and standard channels.→ 140
2. Remove the bottles including the reagent hoses and the bottle tray.
Removing the dispensers
1. Open the dispenser holder by pressing both snap-fit hooks together and removing it.
2. Remove the adapter and the dispenser from the dispenser drive. To do so, hold the
dispenser by the black block at the lower end and pull the block from the metal pin.
Inserting new
1. Make sure that the dispenser is connected correctly.
2. Slide the dispenser holder over the snap-fit hooks. Ensure that the holder snaps into
place.
3. Place the bottle tray back into the housing.
4. Connect the reagent hoses to the corresponding reagent valves.
5. Select
Menu
/
Operation
/
Dispenser replacement
/
Dispenser selection
.
6. Select the dispensers that you have replaced and press
OK
to confirm.
7. Select
Reset operating time counter
.
8. Select
Menu
/
Operation
/
Maintenance
/
Bottle change mode
/
Bottle insertion
/
Bottle selection
.
9. Select all the bottles and press
OK
to confirm.
10. Select
Confirm bottles inserted
.
11. The system must be calibrated after a replacement routine. Select
Menu
/
Operation
/
Manual operation
/
Determine calibration factor
.
12. After the calibration go back to
MODE
/
Continue automatic mode
or
MODE
/
Start
automatic mode
.
12.3 Decommissioning
L
CAUTION
Activities while the analyzer is in operation
Risk of injury and infection from medium!
‣
Before you release any hoses, make sure that no actions, such as the pumping of
sample, are currently running or are due to start shortly.
‣
Wear protective clothing, goggles and gloves or take other suitable measures to protect
yourself.
‣
Wipe up any spilt reagent with a disposable tissue and rinse with clear water. Then dry
the cleaned areas with a cloth.
NOTICE
Do not operate the analyzer for longer than 2 days without taking it out of service
properly!
Can result in damage to the device.
‣
Take the analyzer out of service as described.