Liquiline System CA80SI
Operation
Hauser
67
3. Enter the new password again.
The password has been changed successfully.
Press the navigator for longer to return to the measuring mode.
10.3 Analyzer
Menu/Setup/Analyzer
Function
Options
Info
Mode
Read only
Manual, automatic or fieldbus
Device tag
Customized text,
32 characters
Factory setting
Analyzer_serial number
‣
Select any name for your analyzer. Use the
TAG name for example.
Measuring parameter
Read only
Measuring range
Read only
10.3.1 Extended setup
Measuring value
Menu/Setup/Analyzer/Extended setup/Measuring value
Function
Options
Info
Main value
Selection
• SiO2
• SiO2-Si
• H2SiO3
Factory setting
SiO2
Unit
Selection
• mg/l
• µg/l
• ppb
Factory setting
µg/l
Main value format
Selection
• #
• #.#
• #.##
• #.###
• Auto
Factory setting
Auto
Status for failed measurement
Selection
• Keep last value
• Status bad
Factory setting
Keep last value
Keep last value
Standard practice is to keep the measured value
of the last successful measurement until the next
successful measurement.
To detect an error during measurement, the
NAMUR status signal, the time stamp of the
measurement and the measured value status
must be evaluated.
Status bad
If a measurement cannot be performed due to an
error, the measured value status is set to
Bad
and the measured value to 0. The time stamp is
updated accordingly.