Liquiline System CA80SI
Commissioning
Hauser
45
W
RB
RK
RN
H
SPx
D
SP4
SP3
SP2
SP1
1
2
3
MST
F
S1
RN RK RB
SP5 SP6
FR
FI
RB
RK
RN
S1
SPx
A0036791
61
Hose connection diagram for 4-/6-channel version
D
Sample outlet
RB..N
Reagents RB, RK, RN
F
Flow sensor
S1
Standard 1
FR
Pressure relief valve
SP1..6
Sample inlets
FI
Filter
W
Outlet
H
Heater
1, 2, 3
Dispensers
MST
Magnetic stirrer
9.1.3
Connecting the sample outlet hose "D"
The liquid from sample outlet hose "D" contains only sample mixture. It can be
disposed of accordingly.
Ensure that it can drain freely: Route sample outlet hose "D" without counterpressure.
1-/2-channel version
1. Guide sample outlet hose "D" out of the housing via a hose gland.
2. Attach sample outlet hose "D" to the outlet of the sample channel switching and
secure using a PG coupling with the appropriate clamping unit.
4-/6-channel version
‣
Attach sample outlet hose "D" to the outlet of the panel with the sample channel
switching.