Diagnostics and troubleshooting
Liquiline System CA80SI
124
Hauser
No.
Message
Factory settings
Tests or remedial action
S
D
F
503
Language change
M
On
Off
Language change failed
‣
Contact Hauser Service.
529
Diag. setup active
C
Off
Off
‣
Wait for maintenance to be finished.
530
Logbook at 80%
M
On
Off
1.
Save the logbook to the SD card and then
delete the logbook in the device.
2.
Set memory to circular buffer.
3.
Deactivate logbook.
531
Logbook full
M
On
Off
532
License error
M
On
Off
‣
Contact Hauser Service.
540
Parameter save fail
M
On
Off
Storage of configuration has failed
‣
Repeat.
541
Parameter load ok
M
On
Off
Configuration successfully loaded
542
Parameter load fail
M
On
Off
Loading of configuration has failed
‣
Repeat.
543
Parameter load abort
M
On
Off
Configuration loading aborted
544
Parameter reset ok
M
On
Off
Factory default successful
545
Parameter reset fail
M
On
Off
Setting of device configuration to factory setting
has failed
565
Configuration
M
On
Off
Invalid configuration of sample preparation
system
1.
In
Setup
/
Sample preparation
, check the
number of channels used, their operating
mode and type of installation.
2.
Check the permitted combinations of
sample preparation and analyzers, see the
Operating Instructions for the sample
preparation system.
714
Filter mats change
M
On
Off
Filter mat change necessary
The limit value for operating hours has been
exceeded
‣
Replace the filter mats and reset the
operating hours counter in the Diagnostics
menu.
715
Calibration expired
M
On
Off
The validity of the last calibration has expired.
The date of the last calibration is too far in the
past. Measurement can still take place.
Possible reasons:
Manual intervention has prevented automatic
calibration
1.
Calibrate analyzer manually.
2.
Check device configuration.
716
Calibration expired
S
On
Off
Calibration has failed or is not reliable
Possible reasons:
Stability criteria are not met
1.
Check the configuration and perform a
manual calibration again.
2.
Contact Hauser Service.
717
Photometer defective
F
On
On
Photometer defective
Possible reasons:
• No lamp voltage
• No lamp power
‣
Contact Hauser Service.