Liquiline System CA80SI
Operation
Hauser
69
Menu/Setup/Analyzer/Extended setup/Diagnostics settings/Bottles
Function
Options
Info
Control
Selection
• Off
• On
Factory setting
Off
On
After the monitoring function is switched on,
you can find 2 additional menu items in:
Menu
/
Operation
/
Maintenance
/
Change bottle
• Remaining operating time
• Bottle filling levels
Bottle filling levels
Start volume
Reagent RB
100 to 2500 ml
Factory setting
1000 ml
Reagent RK
100 to 2500 ml
Factory setting
1000 ml
Reagent RN
100 to 2500 ml
Factory setting
1000 ml
Standard S1
100 to 5000 ml
Factory setting
2500 ml
Warning limits
Diag. code 726
Warning limits are intended as an early warning
for maintenance tasks that will need to be
performed soon.
1.
Procure spare parts in time.
2.
Schedule maintenance tasks.
Reagent RB
1 to 40 %
Factory setting
10 %
Reagent RK
1 to 40 %
Factory setting
10 %
Reagent RN
1 to 40 %
Factory setting
10 %
Standard S1
1 to 20 %
Factory setting
5 %
Alarm limits
Diag. code 727
Alarm limits are intended to immediately trigger
maintenance tasks.
‣
Replace the materials concerned as quickly
as possible.
Reagent RB
1 to 40 %
Factory setting
5 %
Reagent RK
1 to 40 %
Factory setting
5 %