Liquiline System CA80SI
Maintenance
Hauser
141
3.
Lift the rubber cover off the cuvette. There must be no kinks in the capillaries at the
capillary holder or at the valves, and the capillaries must not be removed from the
hose connector.
4. Insert the maintenance dispenser together with the hose piece provided through the
hole in the capillary holder of the cuvette.
5. Empty the cuvette completely.
6.
Remove outlet hose "W".
7.
Push the cuvette out from below by pressing simultaneously against the tab and the
hose attachment.