Technical data
Liquiline System CA80SI
160
Hauser
Extension module
Switching voltage
Load (max.)
Switching cycles (min.)
230 V AC, cosΦ = 0.8 to 1
0.1 A
700,000
0.5 A
450,000
2 A
120,000
115 V AC, cosΦ = 0.8 to 1
0.1 A
1,000,000
0.5 A
650,000
2 A
170,000
24 V DC, L/R = 0 to 1 ms
0.1 A
500,000
0.5 A
350,000
2 A
150,000
Minimum load (typical)
• Min. 100 mA at 5 V DC
• Min. 1 mA at 24 V DC
• Min. 5 mA at 24 V AC
• Min. 1 mA at 230 V AC
15.5 Protocol-specific data
PROFIBUS DP
Manufacturer ID
11
h
Device type
155E
h
Profile version
3.02
Device database files (GSD files)
www.endress.com/profibus
Device Integration Manager DIM
Output variables
16 AI blocks, 8 DI blocks
Input variables
4 AO blocks, 8 DO blocks
Supported features
• 1 MSCY0 connection (cyclical communication, master class 1
to slave)
• 1 MSAC1 connection (acyclical communication, master class 1
to slave)
• 2 MSAC2 connections (acyclical communication, master class
2 to slave)
• Device lock: The device can be locked using the hardware or
software.
• Addressing using DIL switches or software
• GSD, PDM DD, DTM
Modbus RS485
Protocol
RTU/ASCII
Function codes
03, 04, 06, 08, 16, 23
Broadcast support for function codes
06, 16, 23
Output data
16 measured values (value, unit, status), 8 digital values (value,
status)
Input data
4 setpoints (value, unit, status), 8 digital values (value, status),
diagnostic information
Supported features
Address can be configured using switch or software