multi N/C 2100S (duo, pharma)
Maintenance and care
83
}
Pull the plug of the Peltier cooling block out of the connection on the rear
wall (see arrow).
}
Take the carrier plate of the TIC condensate container and the condensa-
tion coil out of the mounting bracket on the right side of the furnace.
}
Pull hoses 1, 16 and 21 with FAST connectors off of the connections of
the TIC condensate container and the condensation coil.
}
Pull the condensation coil out of the clamps on the carrier plate (arrow)
and set it aside safely.