Specifications
multi N/C 2100S (duo, pharma)
108
10
Specifications
10.1 Technical data
General characteristics
Designation/type
multi N/C 2100S
multi N/C 2100S pharma
multi N/C 2100S duo
Basic device dimensions (W x H x D)
513 x 464 x 550 mm
Basic device mass
30 kg
Methods data
Digestion principle
Thermocatalytic oxidation
Digestion temperature
Up to 950 °C, depending on catalyst
Measuring methods
TC, TIC, TOC (differential method), NPOC,
DOC, TN
b
Sample feed
Direct injection via septum-free lock
Sample volume
50 to 500 µl
Particle handling capacity
In accordance with DIN EN 1484
Carbon detection principle
NDIR (coupled with the VITA method)
TC, TOC, NPOC, TIC measurement range
0 to 30000 mg/l
TC, TOC in solid measurement range
(with the HT 1300 solids module)
0 to 500 mg
Nitrogen detection
Nitrogen detection principle (optional)
CLD
ChD
TN
b
measurement range (CLD)
0 to 200 mg/l
TN
b
measurement range (ChD)
0 to 100 mg/l
Process control
Control and analysis software
multiWin
Software function scope
Real-time graphics, status indication during
analysis, graphical display of the measured
results, result print-out
Data integrity and conformity with pharma-
ceutical guidelines 21 CFR Part 11 and Eu-
draLex Volume 4 Annex 11 in the pharma-
ceutical software.
Gas supply
Option 1
Oxygen
≥4.5
Option 2
Synthetic air
(from a compressed gas cylinder)
Hydrocarbon and CO
2
-free
Option 3
Purified compressed air
(provided by a TOC gas generator)
CO
2
<1 ppm
Hydrocarbons (as CH
4
) <0.5 ppm
Inlet pressure
400 to 600 kPa
Flow rate
15 l/h, depending on measuring mode