Installation and commissioning
multi N/C 2100S (duo, pharma)
46
CAUTION
Risk of poisoning due to ozone
The ozone generator contained in the device produces ozone (O
3
). When used in accor-
dance with the intended use, the downstream ozone decomposer decomposes the poi-
sonous gas. Various safety measures result in the automatic shut-down of the ozone
generator. The following nevertheless applies:
¡
If there is a sharp smell of ozone, switch the device off immediately and inform cus-
tomer service.
¡
To guarantee perfect and safe operation, Analytik Jena recommends annual inspec-
tion and maintenance by customer service.
Installation on the analyzer
}
Set up the detector next to the analyzer.
}
Connect the carrier gas to the gas connection with quick-release coupling.
}
Set up the gas connection between the detector and the analyzer:
– "sample in" connection on the detector
– "CLD/pump" connection on the analyzer
}
Connect the "CLD/HT" interface on the rear of the analyzer with the RS 232 interface
on the detector via the supplied serial data cable.
}
Switch on the detector. The status LED indicates operational readiness.
Checking and changing the
configuration
}
Switch on the analyzer and the detector. Start the multiWin program and initialize
the analysis system.
}
Check the device configuration with the Instrument | System-Information menu op-
tion in the multiWin set-up info window.
}
If necessary, change the configuration:
– Exit the multiWin program.
– Start the setup tool in the Windows interface with Start | Program files | multi-
Win | multiWin setup tool.
– Select the detector (CLD-TOC) in the N-measurement list.
– Exit the multiWin set-up tool window with [Create].
– Start the multiWin program and initialize the analysis system.
– Open the Options window with menu option Configuration | Edit options. Go to
the Analyzer components tab.
}
In the Sensors group, activate nitrogen measurement with the N-measurement ak-
tive option.
}
Exit the window with [OK].