Troubleshooting
multi N/C 2100S (duo, pharma)
94
Error code: Error message
VERS1: Communication error – analyzer
Cause
Remedy
¡
The analyzer is not switched on.
¡
Switch on the analyzer.
¡
The multiWin software was started too
early.
¡
Only start the software after 30 s.
¡
The analyzer is not connected to the PC.
¡
Check the connection of the analyzer to
the PC.
¡
The incorrect COM port is being used on
the external computer.
¡
Check the port on the external com-
puter.
¡
If necessary, select another interface in
the software (with the Configuration |
Interface menu option).
Error code: Error message
-6: Analyzer is busy
Cause
Remedy
¡
The analyzer is in busy state for
>10 min.
¡
Initialize the analyzer.
Error code: Error message
-5: Communication error – analyzer STAT.
MESS. STEP or INIT
Cause
Remedy
¡
Communication error.
¡
Initialize the analyzer.
Error code: Error message
-4: Communication error – analyzer
Cause
Remedy
¡
Communication error.
¡
Check the interface cable.
¡
Initialize the analyzer.
Error code: Error message
-3: Command from the analyzer CRC error
-2: CRC error
-1: Invalid command from the analyzer
Cause
Remedy
¡
Communication error.
¡
Initialize the analyzer.
Error code: Error message
1: Incomplete command from the PC
2: PC command without STX
3: PC command without *
4: PC command CRC error
5: PC command invalid command
6: PC command invalid MESS command
Cause
Remedy
¡
Faulty connection between the internal
and external program.
¡
Initialize the analyzer.
Error code: Error message
7: COM 2 not found
8: COM 3 not found
9: COM 4 not found
Cause
Remedy
¡
Internal hardware problems.
¡
Switch the analyzer off/on.