multi N/C 2100S (duo, pharma)
Troubleshooting
103
Error
Status LED on the analyzer not lighting
(Lockin)
Cause
Remedy
¡
The internal program has not been
started.
¡
Switch the analyzer off and on again.
Error
Heating monitoring lamp on LED strip not
lighting
Cause
Remedy
¡
Incorrect temperature settings in the
software.
¡
Check the temperature settings under
Configuration |Options in the Analyzer
components tab.
¡
Faulty thermocouple (furnace). The
"Broken Thermocouple" lamp is lit on the
LED strip.
¡
Inform the service.
¡
Faulty electronics component.
¡
Inform the service.
¡
The combustion furnace is not con-
nected correctly.
¡
Check the connection of the combustion
furnace.