Function and design
multi N/C 2100S (duo, pharma)
18
Switch positions:
¡
TC lock closed: Switch set toward the front
¡
TC lock opened: Switch set toward the rear
TC lock closed
TC lock opened
Figure 4 Toggle switch for manual operation of the TC lock
Microliter syringes
Samples are supplied with microliter syringes. The injection volume is 50 to 500 µl. Op-
timum measurement results are reached when 50 to 100 % of the volume of the micro-
liter syringe is used. Various syringes are available. The canulas can be replaced.
For the analysis of particulate samples, the use of canulas with a larger inside diameter
(particle canula) is recommended.
Special microliter syringes with special geometry and a gas connection for NPOC analy-
ses are used for the autosamplers. The syringes have no graduation and are therefore
not suitable for manual operation.
3.1.2
Hose system
Hose diagram
The connection between the individual components is made with labeled hoses. The
numbers and letters circled in the hose diagram correspond to the labels on the hoses in
the analyzer.
/Air
Gas box + pneumatic unit
Needle valve
5/2-way valve
NPOC flow monitoring
NPOC connection
Sample
Cylinder
TC lock
Furnace
Water traps
TIC condensate container
+ dryer
TIC lock
"waste" connection
Condensate pump
Phosphoric acid pump
Water traps
Halogen trap
Flow monitoring
Phosphoric acid
"CLD/pump" connection
NDIR detector
"analyte" connection
"internal" connection
11
12
15
5
6
1
2
21
16
17
22
3
14
19
18
14
13
O
2
Figure 5
Hose diagram