Operation
multi N/C 2100S (duo, pharma)
60
5.4.5
Managing calibration data
Printing calibration data
}
In the Calibration - Calibration Settings window, activate the Use calibration option.
}
Define what to print in the Print options menu:
– Print the calibration graph and/or
– Print individual integrals for each calibrated channel
}
Start printout with the CalibrationReport |Print menu option.
Exporting calibration files
Calibration data is exported via the Data export menu in the Calibration - Calibration
Settings window. You have the following options for exporting calibration data:
¡
Calibration report to an export file
The calibration report (with the *.ajc file extension) is saved to the ...\Calibration ex-
port directory.
¡
Export to a CSV file (with the *.csv file extension)
The CSV file is stored in the preconfigured directory (default ...\multiWin\CSV).
The directory is selected in the Options window in the Files and Directories tab.
Open the window with the Configuration |Edit options menu option.
¡
Export to clipboard
Reopening a calibration report
}
Click on the Data Evaluation |CalibrationReport menu option in the main screen.
}
Select the calibration report in the Selection CalibrationReport window.
If necessary: Set filters in the Selection CalibrationReport window. Or sort the data
records by clicking on a header.
}
Mark the calibration report and click on the [OK] button.
ü
The calibration report is displayed.
5.5
Performing measurements
Note: In the multiWin pharma software, only methods with signing status authorized
can be used for measurement.
5.5.1
Measurement with manual sample feed
}
Create a new method with the Method |New menu option.
}
Alternatively: Load an existing method. Open the Method selection window with the
Method |Load menu option. Mark the desired method. Confirm selection by clicking
[OK].
}
Select manual sample supply by clicking [manual] in the System state window.
ü
The software initializes the analyzer.
}
Check the following displays in the System state window:
– Optical bank
– Nitrogen detector
– Gas flow
– Temperature
}
If a display is red: Check for errors.