Maintenance and care
81
}
Remove the four knurled head screws on the mounting plates of the fur-
nace.
}
Lift the furnace out of the analyzer.
See also
2
Removing the combustion tube [
2
Cleaning the TIC condensate container and condensation coil [
6.8.2
Installing the combustion furnace
}
Open the left side wall of the analyzer. Remove the top cover.
}
Place the furnace on the mounting plates and fasten it with the four
knurled head screws. Tighten the knurled head screws finger-tight.
}
Plug the plug-in connector for the combustion furnace into the socket at
the bottom right of the rear device wall.
}
Install the combustion furnace.
}
Install the TIC condensate container and the condensation coil.
}
Attach the top cover.
}
Close the side wall.
– Connect the protective grounding to the left side wall.
– Slightly tighten the screws first on the bottom side and then on the
top side. Tighten the screws in turns.
}
Open the gas supply. Connect the power plug with the socket and switch
on the analyzer via the main switch.
}
Check the system for leaks.
ü
The analyzer is ready for operation again.