multi N/C 2100S (duo, pharma)
Installation and commissioning
43
Commissioning the autosam-
pler
Figure 21Fasten the autosampler on the analyzer.
}
Switch off the analyzer before installing the autosampler.
}
Place the autosampler on the analyzer.
}
Fasten the autosampler to the housing of the analyzer with the supplied hexagon
socket screws.
}
Connect the low voltage side cable of the table power supply unit to the rear of the
autosampler. Do not connect the power supply unit to the mains power supply yet.
}
Connect the supplied serial data cable to the "sampler" interface on the rear of the
analyzer. Connect the other end of the data cable to the interface on the autosam-
pler.
}
Plug the grounding conductor into the connection on the rear of the analyzer.
}
Connect the supplied waste hose to the wash cup of the autosampler and to the
waste container of the analyzer.
NOTICE! Route the outlet hose at a constant downward incline. If necessary,
shorten the hose. The hose must not dip into the liquid.
}
Place the sample tray and the wash cup on the autosampler.
Inserting the syringe
Figure 22Inserting the syringe
1 Locking screw
2 Clip
3 Septum: NPOC hose connection
4 Locking lever
5 Syringe cylinder
6 Septum: TC lock seal during injections
}
Remove the syringe (without graduation, with connection for NPOC gas) from the
packaging.
}
Connect the syringe to the NPOC hose (not on the multi N/C 2100S pharma model).