multi N/C 2100S (duo, pharma)
Operation
53
5.2
Switching on the analyzer
NOTICE
Risk of device damage due to depleted copper wool
Damage to optical and electronic components of the analyzer due to aggressive combus-
tion products when the copper wool in the halogen trap is depleted!
¡
Only use the device with an operational halogen trap!
¡
Replace the complete filling of the halogen trap when half of the copper wool or
brass wool is discolored!
Before switching on the analyzer, check the following:
¡
The waste hose is connected to a suitable waste container or drain. Free flow is en-
sured. The capacity of the waste container is sufficient.
¡
The gas supply is connected in accordance with regulations and the inlet pressure is
400 to 600 kPa.
¡
There is enough phosphoric acid in the reagent bottle. A volume of 0.5 ml acid is re-
quired per TIC determination.
¡
The halogen trap is connected, filled with copper and brass wool. The copper and
brass wool not used up.
¡
All hoses are properly connected and in good working order.
¡
All optional accessories (autosamplers, solids modules, etc.) are connected.
Prepare the sample and switch on the analyzer as follows:
}
Open the valve on the pressure reducer of the gas supply.
}
Switch on the PC.
}
Switch on the other optional components.
}
Switch on the analyzer via the main switch. The status LED on the left front door
lights up in green.
}
Start the control and analysis software on the PC. Log in with username and pass-
word.
}
Answer the query Initialize analyzer? with [Yes].
ü
After successful login, the software initializes the analysis system and queries the
connected components. The measuring gas flow reaches its target value (160 ml/
min) after approximately 1 to 2 min.
}
If the analyzer is not ready for measurement after 30 min and one or more compo-
nents are still displayed in red in the System state window:
– Check the hose connections for proper fit and check for errors.
}
After each syringe change and after any manipulation of the locks such as changing
the catalyst, check the adjustment of the autosampler. If necessary, readjust the au-
tosampler.
}
After each syringe change: Enter the syringe size in the software.
– Open the Options, window, tab Analyzer components with menu option Configu-
ration |Edit options.
– Select the syringe volume in the Size list field.
}
For NPOC measurements: Set the NPOC purge flow.
The NPOC purge flow has been preset to 50 to 160 ml/min. The purge flow can be
increased or decreased dependent on the measuring task.