Operation
multi N/C 2100S (duo, pharma)
54
ü
The analyzer is ready to measure.
See also
2
2
5.3
Switching off the analyzer
Switching to standby mode
Recommendation: Switch the analyzer to standby mode during measurement pauses of
more than 30 min.
}
Click on [Exit] in the software interface.
ü
The Program End window opens.
}
For measurements with autosampler: Activate the Rinse syringe checkbox. The sy-
ringe is flushed with the solution from the wash cup.
}
Activate the Stand-by Analyzer option and confirm with [OK].
ü
The analyzer remains in standby mode.
The syringe is flushed if syringe flushing is activated. The gas flow is switched off and
the standby temperature is set.
Switching off the device before
extended standstill periods
Recommendation: Switch off the analyzer before extended standstill periods (weekends,
holidays).
}
Click on [Exit] in the software interface.
ü
The Program End window opens.
}
For measurements with autosampler: Activate the Rinse syringe checkbox. The sy-
ringe is flushed with the solution from the wash cup.
}
Activate the Switch off Analyzer option and confirm with [OK].
ü
The analyzer is switched off.
If backflushing is enabled, the syringe is flushed. The TIC condensate container is
pumped out. The analyzer shuts down. The gas flow is switched off and the furnace
cools down.
5.4
Carrying out calibration
5.4.1
Preparing and starting calibration
The control and analysis software provides the option to adjust the analysis to the indi-
vidual measuring task by selecting the method. An ideal measurement with its corre-
sponding method requires its own calibration for each analysis parameter and each
measuring channel. Not all parameters in a method need necessarily be calibrated.
Three calibration functions can be stored for each parameter in a method.
Carry out the calibration as follows:
}
Select the sample feed type in the System state window:
[manual] or [Sampler].
ü
The analyzer is initialized.