multi N/C 2100S (duo, pharma)
Maintenance and care
91
6.12 Removing the integrated solids module
CAUTION
Risk of burns at the hot furnace
¡
Switch off the device and allow it to cool before performing maintenance.
}
Exit the software.
}
Switch off the analyzer via the main switch and disconnect the power
plug from the socket. Cut the gas supply.
}
Disconnect the plug-in connector on the left side of the module.
}
Disconnect the carrier gas hose from the FAST connector on the carrier
gas outlet of the gas box. Remove the angled FAST connector from the
gas box connection. Attach the straight FAST connector.
}
Unscrew the other hose end from the solids module.
}
Remove the fork clamp at the spherical joint between the measuring gas
hose and the condensation coil inlet.
}
Loosen the four knurled head screws on the holding plate and pull the
module out of the combustion furnace.
The measuring gas hose and the holding plate can remain on the mod-
ule. This makes the next installation easier.
NOTICE! Do not remove the angled profiles from the furnace. These
profile have been preadjusted and ensure the correct installation posi-
tion
.