multi N/C 2100S (duo, pharma)
Maintenance and care
79
}
Place the pressure ring (2) in the union nut.
The conical side of the pressure ring must point upward.
}
Slide the three coated sealing rings (3) onto the combustion tube.
Ensure that the sealing rings are flush at the edge of the combustion
tube.
}
If it is not present, insert the ceramic holder into the top opening of the
combustion furnace.
}
Insert the combustion tube into the combustion furnace.
}
Screw the lock loosely into the holder with the hexagon socket screw.
}
Hold onto the combustion tube at the bottom. Carefully put the TC lock
onto the combustion tube up to the stop.
}
Press the lock lightly against the combustion tube and screw in the union
nut hand-tight.
}
Screw the lock into the holder with the hexagon socket screw.
}
Screw the Fingertight connection of the carrier gas connection to the TC
lock.
}
Place the top cover on top of the analyzer.