Maintenance and care
multi N/C 2100S (duo, pharma)
92
}
Remove the sealing plug from the vertical opening of the combustion fur-
nace. Insert the plug into the horizontal opening of the combustion fur-
nace.
}
Reinstall the combustion tube for vertical operation.
See also
2
Installing the combustion tube [
6.13 Maintaining the chemiluminescence detector (CLD)
Replace the adsorber cartridge on the rear of the detector every 12 months. The car-
tridge cleans the gas which exits the detector at the "out" outlet.
The cartridge is filled with active carbon and soda lime. Do not open the cartridge. Dis-
pose of the used cartridge as a whole in accordance with the local regulations.
Figure 41Replacing the adsorber cartridge
}
Disconnect the hose from the cartridge.
}
Pull the cartridge out of the clamp.
}
Unscrew the hose connection at the top of the cartridge.
}
Dispose of the used cartridge as a whole in a professional manner.
}
Screw the hose connection into the top of the new cartridge.
}
Push the new cartridge into the clamp. Connect the cartridge to the hose from the
"out" outlet.
ü
The detector can now be used for measurements again.