multi N/C 2100S (duo, pharma)
Security
11
Mandatory action
labels/information
symbols
Meaning
Comment
Observe the operating
manual
On the side parts and the rear of the
device: Before starting work, read the
user manual.
For People's Republic of
China only
The device contains controlled sub-
stances. Analytik Jena GmbH warrants
that these substances will not be re-
leased from the device within the next
25 years provided the device is used as
intended.
2.2
Requirements for the operating personnel
The device must only be operated by qualified specialist personnel instructed in the
use of the device. This instruction also include teaching the contents of this user
manual and of the user manuals of the connected system components. We recom-
mend training by qualified employees of Analytik Jena or its representatives.
In addition to the safety instructions in this user manual, the general applicable
safety and accident prevention regulations of the respective country the device is op-
erated in must be observed and adhered to. The operator must ensure the latest ver-
sion of these regulations.
The user manual must be accessible to the operating and service personnel.
2.3
Safety instructions, transport and commissioning
Incorrect installation can create serious hazards. This may result in electric shock and
explosion if the gases are not connected correctly.
¡
Only the Analytik Jena GmbH customer service or specialist personnel trained
and authorized by them is allowed to install and commission the device and its
system components.
¡
Unauthorized assembly and installation is not permitted.
Insufficiently secured components pose a risk of injury.
¡
During transport, secure the device components as specified in these operating
instructions.
¡
Loose parts must be removed from the system components and packed sepa-
rately.
To prevent health damage, the following must be observed when moving the device
in the laboratory (lifting and carrying):
¡
For safety reasons, two persons are required to transport the device who must
hold the unit by either side of the equipment.
¡
The device does not have any carrying handles. Therefore the device must be
gripped firmly with both hands at the lower end.
¡
Risk of damage to health due to improper decontamination! Perform a profes-
sional and documented decontamination of the device before returning it to An-
alytik Jena. The decontamination report is available from Service when register-
ing the return. Service must refuse acceptance of contaminated devices. The
sender may be liable for damage caused by inadequate decontamination of the
device.