multi N/C 2100S (duo, pharma)
Transport and storage
105
9
Transport and storage
9.1
Transport
When transporting the device, observe the safety instructions in the "Safety instructions"
section.
Avoid the following during transport:
¡
Impact and vibration
Risk of damage due to shock, impact or vibration!
¡
Large temperature fluctuations
Risk of condensation!
9.1.1
Preparing the analyzer for transport
CAUTION
Risk of burns from the furnace
The combustion furnace is still hot after the device has been switched off. There is a risk
of burns.
¡
Allow the device to cool before removing the combustion furnace.
CAUTION
Risk of injury
A risk of injury due to broken glass is present when handling glass parts.
¡
Handle glass parts with extreme caution.
NOTICE
Risk of device damage due to unsuitable packaging material
¡
Only transport the device and its components in the original packaging.
¡
Empty the device completely and attach all transport locks before transporting the
device.
¡
Add a suitable desiccant to the packaging to prevent damage from moisture.
Prepare the analyzer for transport as follows:
}
Rinse the phosphoric acid pump and the corresponding hoses with ultrapure water.
Empty the components.
}
Switch off the analyzer via the main switch. Allow the device to cool down.
}
Cut the gas supply. Disconnect the power plug from the power socket.
}
Disconnect all cables and gas hoses on the rear of the analyzer.
}
Open the doors of the analyzer.