multi N/C 2100S (duo, pharma)
Troubleshooting
97
Error code: Error message
200: Restart computer in the analyzer
Cause
Remedy
¡
Internal computer reset.
¡
Overvoltage.
¡
Short-term power failure.
¡
If the LED status display is lit, initialize
the analyzer.
¡
For repeated occurrences, monitor pre-
cisely at which time the error occurs. Ob-
serve the status bar.
Error code: Error message
201: Restart the internal program
Cause
Remedy
¡
Internal program error.
¡
Initialize the analyzer.
¡
For repeated occurrences, monitor pre-
cisely at which time the error occurs. Ob-
serve the status bar!
Error code: Error message
MESSx: Analyzer error: MESSx measurement
is cancelled
Cause
Remedy
¡
Device error.
¡
Initialize the analyzer.
¡
Acknowledge the error. Observe the dis-
play in the System state window.
¡
Search for the cause and remedy the
fault.
Error code: Error message
Peltier temperature outside range
Cause
Remedy
¡
Peltier cooling insufficient.
¡
Inform the service.
¡
Replace the water traps after repair.
Error code: Error message
Minimum sample volume > cup volume
Cause
Remedy
¡
For sample supply via autosampler:
Selected sample volume too large.
Number of determinations too high.
¡
Check the following settings in the
method: Sample volume/purge volume.
¡
Adjust the number of determinations
(repeat measurements) to the vessel
volume.
7.2
Status errors
Status errors are displayed in red or yellow in the System state window.
Error indication
In 160 ml/min; Out < 150 ml/min
Cause
Remedy
¡
The union nut on the combustion tube is
not tightened correctly (after catalyst re-
placement).
¡
The carrier gas supply to the furnace
head or lock is not connected properly
(after catalyst replacement).
¡
The sealing rings on the combustion
tube are defective (heavily deformed) or
not attached (after catalyst replace-
ment).
¡
The FAST connector on the TIC conden-
sate container is leaky.
¡
Check the screw connections for com-
pleteness and deformations. Tighten as
necessary.
¡
Check the carrier gas supply, in particu-
lar the FAST connector on the analyzer
wall and the screw connection on the
furnace head.
¡
Check all connection points on the water
traps. Replace the FAST connector as
necessary.