Maintenance and care
multi N/C 2100S (duo, pharma)
84
}
Remove the four screws on the side holding the Peltier cooling block on
the insertion tray for the TIC container.
}
Remove the TIC condensate container from the tray. Carefully pour the
acidic solution into a beaker.
}
Check the TIC condensate container and condensation coil for deposits
and cracks.
}
Rinse both glass parts with ultrapure water and dry it well.
Installation
}
Slide the rubber ring onto the bottom adapter of the condensate con-
tainer. The ring protects the glass vessel from the metal holder.
}
Place the TIC condensate container in the tray of the carrier plate.
}
Screw the side of the Peltier cooling block to the tray with four screws.