multi N/C 2100S (duo, pharma)
Maintenance and care
89
– The labeling on the small water trap (disposable retention filter) must face down-
ward.
}
Connect the large water trap with the upper FAST connector.
}
Connect the small water trap to the Luer connection on the bottom.
}
Press the water traps into the clamps on the gas box.
}
Close the side wall.
– Connect the protective grounding to the left side wall.
– Slightly tighten the screws first on the bottom side and then on the top side.
Tighten the screws in turns.
}
Connect the power plug with the socket and switch on the analyzer again via the
main switch.
}
Check the system for leaks.
ü
The water traps on the front side and the gas box are replaced.
See also
2
Checking the system for leaks [
6.11 Replacing the halogen trap
NOTICE
Risk of device damage due to depleted copper wool
Damage to optical and electronic components of the analyzer due to aggressive combus-
tion products when the copper wool in the halogen trap is depleted!
¡
Only use the device with an operational halogen trap!
¡
Replace the complete filling of the halogen trap when half of the copper wool or
brass wool is discolored!
The analyzer can remain switched on to replace the used copper and brass wool.