Maintenance and care
multi N/C 2100S (duo, pharma)
74
Angled FAST connectors
With angled FAST connectors, do not slide the hose ends beyond the side length of the
connector. The gas flow will otherwise be impaired.
Figure 36FAST connector, angled
1 Angled FAST connector
2 Hose
3 Glass connection
Fingertight connections
}
When replacing Fingertight connections, only use straight cut, round, uncrimped
hose ends.
}
Slide the conical nipple onto the hose with the conical side towards the banjo bolt.
The conical nipple and hose end must be flush.
}
Do not jam the banjo bolt during insertion and only tighten it hand-tight.
1
2
3
Figure 37Replacing the Fingertight connection
1 Hose
2 Banjo bolt
3 Conical nipple
6.6
Checking the system for leaks
NOTICE
Risk of gas leakage
When the outlet flow is significantly less than the inlet flow, the device system has a gas
leak.
¡
Check all connection pieces, for example with a foamy tenside solution.
¡
Only put the device into operation when the gas leak has been eliminated.
The system tightness is automatically checked at the gas outlet of the analyzer.