Maintenance and care
multi N/C 2100S (duo, pharma)
82
6.9
Cleaning the TIC condensate container and condensation coil
The TIC condensate container and condensation coil are mounted on a carrier plate on
the right-hand side of the furnace.
Removal and cleaning
CAUTION
Risk of burns at the hot furnace
¡
Switch off the device and allow it to cool before performing maintenance.
}
Switch off the analyzer via the main switch. Disconnect the power plug
from the socket. Shut off the gas supply on the pressure reducer in the
laboratory.
}
Open the left side wall of the analyzer. Push the accessory modules to the
side if necessary. Do not kink any connection hoses.
– Unscrew the four attachment screws. The screws are captive and re-
main attached to the wall.
– Remove the protective grounding. Set the side wall aside safely.
}
Pull hose 2 (to the water traps) off of the top outlet of the TIC conden-
sate container.
}
Release the joint connection (3) on the bottom of the combustion fur-
nace connecting the combustion tube with the condensation coil.
}
To do this, unscrew the knurled head screw (1) and remove the fork
clamp (2).
}
Remove the knurled head screw for attaching the carrier plate.