Analyzer Variables
Rosemount Analytical Inc. A Division of Emerson Process Management
2-32
NGA2000 Reference
LIN_MEASA
LIN_MEASB, LIN_MEASC.
Direction Bidirectional
Binding
LIN_MEASSA, etc.
Type
7-element floating point (except LIN_MEASC, 6-element) array
Value range
0 - 1,000,000
Units ppm
Description
Linearization function measured gas value.
Usage
Used in the "Tweak" and self linearization functions to store the measured gas
concentrations corresponding to the actual concentrations in LIN_GASESA, etc.
See also
LIN_ORDER
Direction Output
Binding
LIN_ORDER
Type Integer
Value range
1-4
Units None
Description
Order of polynomial calculated by self linearization function.
Usage
Shows the order of polynomial calculated.
See also
LIN_OVER
Direction Bidirectional
Binding
None
Type
4-element floating point array
Value range
0 - 100; normally 10
Units Percent
Description
The amount greater than the linearization range over which this curve is valid.
Usage
Linearization curves are valid only over a certain range given by LINRNGHI.
However we allow a certain leeway above this point so that any sudden change of
slope caused by disabling the linearizer does not cause odd effects in the PVA
value. This LIN_OVER variable sets how much leeway there is in percent of the
LINRNGHI value. Above this point the algorithm will provide a linear relationship
between input and output, with a slope equal to the slope of the linearization curve at
the LINRNGHI value.
See also
Other LINxxxxxxxx variable
LIN_RESULTA
LIN_RESULTB, LIN_RESULTC
Direction Bidirectional
Binding
LIN_RESULTA, etc.
Type
7-element floating point (except LIN_RESULTC, 6-element) array
Value range
0 - 1,000,000
Units ppm
Description
Linearization function measured gas value as linearized.
Usage
Used in the "Tweak" and self linearization functions to show the linearized measured
gas concentrations corresponding to the actual concentrations in LIN_GASESA, etc.
See also