Reference Manual
748384-C
September 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Analyzer Variables
2-43
NGA2000 Reference
PELT_OV
Direction Output
Binding
None
Type Enumerated
Value range
NORMAL; NORMAL; WARNING
Units None
Description
CLD only; warning message for the peltier control algorithm menu
Usage
Used to show a problem with the Peltier cooler, the device used to cool the optical
detector in the CLD.
See also
PELT_SET
Direction Output
Binding
None
Type Integer
Value range
Units Volts
Description
CLD only; the voltage being put on the Peltier device
Usage
Shows the voltage with which the Peltier cooler is currently being driven.
See also
PERCENT_MOD
Direction Output
Binding
None
Type
Floating point
Value range
0-100
Units Percent
Description
NDIR only; the modulation percentage resulting from the modulation check function.
Usage
The modulation check function allows the user to determine the amount of
modulation, or change in light transmission, caused by the span gas. There is a
sequence of events that are controlled by a menu in the NDIR analyzer. The result of
this function is reported in this variable. Typical values should be between 10 and
30%.
See also
PFACTOR
Direction Bidirectional
Binding
None
Type
2-element floating point array
Value range
Element 1 - 1250; element 2 - 1.5
Units None
Description
Offset and slope respectively for the pressure compensation algorithm.
Usage
Used to determine how the pressure compensation algorithm works.
See also
BAROMETER
PCOMP_GAIN