Reference Manual
748384-C
September 2003
Analyzer Variables
Rosemount Analytical Inc. A Division of Emerson Process Management
2-54
NGA2000 Reference
ST2NAME
Direction
Analyzer input
Binding
None
Type
Single element enumeration
Value range
Analyzer specific: FID - FLAME OUT, PURGE FAILURE; CLD - OZONATOR OFF,
NO MODE, NOx MODE; others INVALID SELECTION.
Units None
Description
Each bit of the third and fourth elements of STCONT shows the state of the
corresponding alarm named in this variable; these are specific alarms for each
analyzer type.
Usage
Used to allow the 1/0 Module to select which specific alarm to assign to which relay.
These values are shown on the 1/0 module alarm selection menu. INVALID
SELECTION is not shown on the menus.
See also
ST1NAME
ST3NAME
ST4NAME
STCONT
ST3NAME
Direction
Analyzer input
Binding
None
Type
Single element enumeration
Value range
INVALID SELECTION in all single channel analyzers. Same as ST1NA2v1E in MLT
analyzers.
Units None
Description
Provided to allow compatibility with MLT analyzers.
Usage
Not used in analyzers covered by this document. INVALID SELECTION is not shown
on the menus. The fifth and sixth elements of STCONT apply to this variable.
See also
ST1NAME
ST2NAME
ST4NAME
STCONT
ST4NAME
Direction
Analyzer input
Binding
None
Type
Single element enumeration
Value range
INVALID SELECTION in all single channel analyzers. Same as ST1NAME in MLT
analyzers.
Units None
Description
Provided to allow compatibility with MLT analyzers.
Usage
Not used in analyzers covered by this document. INVALID SELECTION is not shown
on the menus. The seventh and eighth elements of STCONT apply to this variable.
See also
ST1NAME
ST2NAME
ST3NAME
STCONT