Reference Manual
748384-C
September 2003
1-28 Introduction
Rosemount Analytical Inc. A Division of Emerson Process Management
NGA2000 Reference
1-13 SECURITY
You can enable security for each or all of the
three main branches of the menus. In each
case, you have to enter a five letter password in
order to enter the appropriate section of the
menus. Once you have entered the password,
you will be able to enter or leave that section of
the menus at will, unless you leave the instru-
ment unused for a time out period, in which
case you will have to re-enter security. If you
make the passwords the same for each branch,
entering one branch will disable security for all
branches.
You can also set a time out for the menus
themselves, after which the display will jump out
of the menus. This in effect is another security
device - it is practically impossible to figure out
what to do in time if you set the time out to 10
seconds, thus it keeps people from making un-
authorized adjustments.
Pressing the "LOCK" button on the main display
screen will enable security immediately, cutting
short the normal hold-open time.
a. Basic
Instructions
Enter the Main menu…
Select Technical level configuration...
Select System setup...
Select Security codes...
Enable whichever set of codes you wish to
use.
Select whichever security code you wish to
configure.
Press the buttons in a suitable order. The
code will be shown as a number.
Press the left arrow key when you are done.
Setup the other security codes as desired.
Press the left arrow key until you are in the
System setup menu.
Select Front panel control...
Set the Drop out of menus after: line to a
suitable time.
This line controls how long it will leave the
menus showing after you have pressed a
key in one.
b. Troubleshooting
Someone put in a password and now I'm
locked out…
Call Rosemount Analytical Service and they
will tell you a secret back door password.
This will allow you to enter the
Technical
level configuration
security screen, and then
you can reset all the other passwords.
I put in passwords but I still don't have
any security screens…
You have to enable security for each level
you want it to work on. Do this in the
screen immediately above the security
password entry screens.