Reference Manual
748384-C
September 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Analyzer Variables
2-19
NGA2000 Reference
CURRENTINT
Direction Output
Binding
None
Type Enumerated
Value range
OFF; ON
Units None
Description
Shows status of interference compensation for this range
Usage
Not used in version 2.2 or earlier. Version 2.2 analyzers do not have the interference
compensation feature.
See also
CURRENTLSTAT
Direction Output
Binding
CURRENTLSTAT
Type Enumerated
Value range
DISABLED; ENABLED – 1; ENABLED - 2; ENABLED - 3; ENABLED - 4.
Units None
Description
Shows status of linearization for the current range.
Usage
Used for displaying the linearization status of the current range. It is possible to use
any of the sets of linearization coefficients on a given range, as long as the upper
limit of the range is within the limit of the linearizer range.
See also
LINSTAT
LINFORRANGE
CURRENTMTHD
Direction Output
Binding
CURRENTMTHD
Type String
Value range
31 characters
Units None
Description
Unused. Was to show the parameter set currently in use, but this function has not
been implemented.
Usage
Not used in version 2.2 or earlier.
See also
CURRENTRNGHI
Direction Output
Binding
None
Type Floating
point
Value range
Same as RNGHI
Units ppm
Description
The value of the upper limit point of the current range.
Usage
Used to show the current range's upper limit point on menus, rather than the entire
array of range end points as would be shown by RNGHI. The analyzer copies the
value of the appropriate element of RNGHI into this variable whenever CRANGE is
changed.
See also
RNGHI
CRANGE