Reference Manual
748384-C
September 2003
Rosemount Analytical Inc. A Division of Emerson Process Management Analyzer Variables
2-61
NGA2000 Reference
VOLTS_IS
Direction Output
Binding
None
Type
5-element (6-element in FED, CLD) floating point array
Value range
Nominal voltage ± about 2V
Units Volts
Description
Reports internal power supply voltages:
+15 V
-15 V
+ 5 V; +10 V FID
+24 V; polarizing voltage (~95 V) FID
+12 V; not on FID On HFID, 10 V reference.
+10 V CLD only
Usage
Reports internal power supply voltages. Note that a significant error in the +5 V supply
will not be reported as it will stop the computer from working.
See also
AMVOLTSWAS
ZERO
Direction Bidirectional
Binding
None
Type
4-element floating point array
Value range
Around 500,000 except CLD, 32,000
Units None
Description
Zero offset. Element 1 refers to the zero offset for range 1, etc.
Usage
Contains the zero offsets used in the main PVA calculation. During zero calibration,
these are adjusted such that the value of PVA resulting is within 0.02% of the full scale
range value of the appropriate element of ZEROGAS. If CALCHCKLIMITS is set to
enabled, this will fail if the change required is greater than the percentage of the full
scale value given in CALFAIL. The value of ZERO should be the same as the value of
RAW-SIGNAL while zero gas is flowing, and a zero may be effected by simply copying
the value of RAW-SIGNAL into the appropriate element of ZERO.
See also
SPAN_GOOD
ZEROGAS
ZERO_WAS
ZERO_COMP
Direction Bidirectional
Binding
None
Type
Floating point
Value range
Around 0
Units None
Description
NDIR and PMD only; determines the temperature compensation for the offset
coefficient.
Usage
Used to set the zero temperature compensation in the NDIR and PMD analyzers.
See also