Analyzer Variables
Rosemount Analytical Inc. A Division of Emerson Process Management
2-30
NGA2000 Reference
IOHOLDCAL
Direction
Analyzer: bidirectional; 1/0 module: output
Binding
IOHOLDCAL
Type Enumerated
Value range
OFF; HOLD; CEMS
Units None
Description
Allows the analyzer to control the track and hold operation of the 1/0 module when
the analyzer module is in a calibration mode
Usage
OFF - the 1/0 module tracks the analyzer reading through a calibration event.
HOLD - the 1/0 module holds the last reading when the analyzer PROCESS variable
is anything other than 0.
CEMS - forces the 1/0 module to track the PVACEMS reading rather than the PVA
reading while PROCESS is greater than 0.
I/0 Module:
If set to DISABLED, the output will track the PVA reading during a
calibration cycle. If set to ENABLED, the output will hold at its last value as soon as a
calibration cycle is entered (or if the analyzer is manually told to do a zero or span). If
set to CEMS, the output tracks PVACEMS (see this variable in the analyzer list)
during a calibration cycle.
See also
PROCESS
CALSTAT
PVACEMS
PVA
IOMAINTHOLD
J13_IN
Direction Output
Binding
None
Type Floating
point
Value range
-5 - +5
Units Volts
Description
CLD only; values of J13 input
Usage
Unused. Shows the value of this input to the secondary ADC.
See also
J6_IN
Direction Output
Binding
None
Type Floating
point
Value range
-5 - +5
Units Volts
Description
CLD only; values of J6 input
Usage
Unused. Shows the value of this input to the secondary ADC.
See also
LASTSPAN
Direction Output
Binding
None
Type Floating
point
Value range
0 - the analyzer PVA limit
Units ppm
Description
The last span gas reading, using the ZERO factor from the subsequent zero
operation.
Usage
Shows any interaction between zero and span operations. If the zero operation
affects the span, this will show a change. It is most useful when tying to calibrate an
analyzer with a non-zero zero gas.
See also
LASTZERO