Reference Manual
748384-C
September 2003
Rosemount Analytical Inc. A Division of Emerson Process Management I/O Modules Variables
3-25
NGA2000 Reference
TAGC
Direction Input
Binding
None
Type String
Value range
31 characters
Units None
Description
SCAL only; shows the AMTAG belonging to analyzer 3.
Usage
Allows the menus for the S CAL module to identify which analyzer is regarded as
analyzer 3.
See also
AMTAG (in Section 2)
TAGA
TAGB
TAGD
TAGD
Direction Input
Binding
None
Type String
Value range
3 1 characters
Units None
Description
SCAL only; shows the AMTAG belonging to analyzer 4.
Usage
Allows the menus for the SCAL module to identify which analyzer is regarded as
analyzer 4.
See also
AMTAG (in Section 2)
TAGA
TAGB
TAGC
TESTMODE
Direction Output
Binding
None
Type
2-element enumerated array
Value range
AUTO; TEST
Units None
Description
Controls the 1/0 algorithm to allow direct user modification of output devices,
The relay control.
The analog output control.
Usage
Setting either element to TEST causes the various variables that can directly set
the output to be operational. The module will reset the variable to AUTO after 1
minute.
See also
RSTATUS
CAL_
OP