Reference Manual
748384-C
September 2003
Analyzer Variables
Rosemount Analytical Inc. A Division of Emerson Process Management
2-22
NGA2000 Reference
DATA_POINT
Direction Output
Binding
DATA_POINT
Type
7-element floating point array
Value range
See description below
Units
Depends on variables selected
Description
Contains the data for the particular 15 minute average segment selected by
DATA_INDEX.
Usage
Set DATA_INDEX to the segment of stored data desired, then after about 1 second
this variable will be loaded with the data appropriate for that segment. If no data is
available for the particular segment, elements 0 and 5 will contain a large negative
number. The elements are as follows:
Primary variable (normally, but selected by TRENDCONT1)
Day of month
Hour
Minute
If > 9, calibration occurred; if 1 or 11, at least one invalid reading
Secondary variable (from TRENDCONT2)
The index where this data segment is stored, 0 - 95
Stored variables are selected by TRENDNAME, which shares data with
TRENDCONT to tell the analyzer which variables to store.
See also
DATA_POINT
TRENDNAME
TRENDCONT
DEBUG
Direction Output
Binding
None
Type
7-element floating point array, 7-element integer array
Value range
Undetermined
Units None
Description
Debug variables used in software development
Usage
Used only for software development, Undetermined contents
See also
DEBUGI
DEBUGI
Direction Output
Binding
None
Type
7-element floating point array, 7-element integer array
Value range
Undetermined
Units None
Description
Debug variables used in software development
Usage
Used only for software development, Undetermined contents
See also
DEBUG