Reference Manual
748384-C
September 2003
Analyzer Variables
Rosemount Analytical Inc. A Division of Emerson Process Management
2-36
NGA2000 Reference
MEASUREPOINT
Direction Bidirectional
Binding
MEASUREPOINT
Type Enumerated
Value range
Point 1; Point 2; Point 3
Units None
Description
Allows the user to specify which of the three piecewise linear approximation points
he is dealing with.
Usage
The "Tweak" function allows the user to adjust up to three midscale points per range
in order to make the linearization curve go exactly through particular standard gases.
See also
METHODA
METHODB, METHODC, METHODD, METHODE
Direction Bidirectional
Binding
None
Type String
Value range
31 characters
Units None
Description
Unused. Intended to be the description of the parameter sets.
Usage Unused.
See also
MID_GASA
MID_GASB
Direction Bidirectional
Binding
MID_GASA, MID_GASB
Type
6-element floating point array
Value range
-1 - 1,000,000
Units ppm
Description
Elements 1, 2 and 3 are the mid gas values as entered for range 1; the next three
points are those for range 2, and so on.
Usage
Allows the user to enter the desired gas values for the three midrange points on
each range. The "Tweak' function can then force the measured values to equal
these, interpolating other points between them.
See also
MIDPOINTA
MIDPOINTB
Direction Bidirectional
Binding
MIDPOINTA, MIDPOINTB
Type
6-element floating point array
Value range
-1 - 1,000,000
Units ppm
Description
Elements 1, 2 and 3 are the mid gas values as read for range 1; the next three points
are those for range 2, and so on.
Usage
See MID_GASA, MID_GASB. These are the measured values for those gases,
before they are forced to be equal to those values by the "Tweak" function.
See also